Tech Tip | Scanner Reset error code, GE Lunar Prodigy

Tech Tip | Scanner Reset error code, GE Lunar Prodigy

Is the Scanner Reset error code an SBC Board error?

CSBC_BOARD

CSBC_BOARD

The Scanner Reset error code is actually one of the more common error codes on the GE Lunar Prodigy bone density systems. At first glance, a technician may diagnose this error as an issue with the SBC Board.

Tech Tip_SBC Board_Error Codes

Tech Tip_SBC Board_Error Codes

However, in most instances, the Scanner Reset error code is actually not related to an SBC Board failure.

Since the SBC Board monitors all the systems, it is understandable that the Scanner Reset error code message may trigger this diagnosis. Always best not to jump to this conclusion right away, and find the correct cause of the Scanner Reset error code.

In order to properly assess the Scanner Reset error code, a technician should first go into the diagnostic testing tab and run a complete diagnostic test – of all the ‘testable parts’. This will help to correctly to determine the ultimate cause of the error.

Find more tech tips here.

For more information on this scanner reset error code, GE Lunar Prodigy diagnosis, or any other parts questions, feel free to contact me personally.

Steve Adams, Parts Manager